The advantages of 2, 3 and 4 wire Pt100 measurement.

INTRODUCTION 
Two, Three and Four wire measurement techniques have been developed for measuring accurately the resistance of resistive temperature detectors (RTD). This application note looks at the new techniques being adopted in ‘Smart’ instruments where, using the intelligence of a microprocessor, the traditional drawbacks of 3 wire systems no longer apply.
THREE WIRE MEASUREMENT 
The traditional method of accurately measuring a resistance, is to incorporate the resistance into a Wheatstone bridge circuit (see figure below).



A voltage excites the bridge and the voltage across the bridge is proportional to the resistance of the RTD.

A problem occurs when we introduce lead resistances (See figure below). It is apparent that any resistance in the lead looks as though there is additional resistance in the element to be measured.





To minimise these errors the three wire compensated bridge was introduced (see figure below).



This has the effect of removing the error introduced by the lead resistance as long as lead resistances RL1 and RL3 are matched.
However the effect of the lead resistance can be to cause less current to flow in the detector leg and hence introduce a small but possibly significant span error. This can be eliminated by exciting the bridge from a constant current source rather than a constant voltage and so whatever the lead resistance, the same current always flows through the detector. With this method there are no lead resistance errors introduced as long as the lead resistances are equally matched. In practice they are very closely matched as long as the wire used is part of the same multi-core cable.
The exception to this is when the sensor is used in a Hazardous area and connected to the bridge circuit via a Zener Barrier. Here any mis-match in the resistance of the two legs of the Zener barrier can appear as a sensor error. Although still small, this error can be as much as 0.15 ohm or approx. 0.3°C.(MTL 155 Barrier).
For analogue transmitters Status Instruments Ltd. have traditionally used a variation of this technique using an in house active bridge circuit. The exception being the new ‘Smart’ series of instruments which use a different technique which will be explained later.

Another way of measuring Pt100 elements is to use a 4 wire current and voltage method (see figure below).



Here the detector is excited by a constant current and the voltage across the detector measured by an amplifier with a high impedance input. If the current source is perfect and the input impedance of the voltage measuring circuit is infinite, then there is no error whatsoever introduced by the lead resistances even if they are mismatched.
THE ‘SMART’ WAY FORWARD 
This paragraph describes the method used to measure Pt100 on the new DM3000 series instruments.
The current trend for so called SMART instruments is to have a universal input capable of supporting a wide range of inputs. It is inconvenient (and unnecessary) to dedicate input pins and electronics to support a constant current supply and a bridge arrangement. The input circuit measures voltages to a high degree of accuracy and the microprocessor performs the calculation in the figure below.



Rc is used solely to limit the current flowing and Rs is a stable reference resistor.
Having computed the resistance, the microprocessor applies the corrections required and translates the resistance to an accurate temperature reading.
In addition, the microprocessor can determine which if any, of the RTD inputs has become disconnected and detect other errors such as RTD short circuit. This is an improvement over both conventional three and four wire circuits because you can now have a predictable failure mode which does not depend upon which of the three wires has become disconnected.
This technique removes lead resistance effects as long as they are equal. Again, we have the problem when using Zener barriers, in that if the legs of the barrier are not accurately matched, then a small error could be introduced.

Comments

Popular posts from this blog

Installing and Activating KEIL uVision

*** NOTE: Click on the images to view them in full resolution Download: Download Keil from the link below: Keil Latest Version Free download with keygen After Installation: 1. File -> License Management 2. Open Keygen from the folder where rar file is unpacked 3. Copy computer ID ( CID ) to clipboard 4. Paste CID in KEYGEN exe, Press generate, copy generated serial and paste in new license textbox of Licence management. *** NOTE: Click on the images to view them in full resolution

Average and Square Root Using 8085

AVERAGE OF N NUMBERS AIM:       To write an assembly language program that finds the average of N numbers and executes the program using 8085 microprocessor kit. APPARATUS REQUIRED: 8085 microprocessor kit, power supply. ALGORITHM: STEP 1: Load HL register pair with a memory pointer. STEP 2: Get the count to B register and clear accumulator. STEP 3: Push the count to stack pointer. STEP 4: Increment memory pointer. STEP 5: Add the content of memory location to accumulator. STEP 6: If carry results increment D register. STEP 7: Decrement count. If it is not zero go to STEP 4. STEP 8: Else move the sum to E register. STEP 9: Get the count from stack pointer. STEP10: Subtract the count from sum(content of DE pair). STEP11:If the subtraction results a carry add the count to the result and get the remainder STEP12: Otherwise increment B register content and go to STEP10. STEP 13: Store the quotient and remainder in successive memory location. PROGRAM:

Understanding Absolute and Kinematic Viscosity

        Of all the tests employed for used oil analysis, none provides better test repeatability or consistency than viscosity. Likewise, there is no property more critical to effective component lubrication than base oil viscosity. However, there is more to viscosity than meets the eye. Viscosity can be measured and reported as dynamic (absolute) viscosity or as kinematic viscosity. The two are easily confused, but are significantly different. Most used oil analysis laboratories measure and report kinematic viscosity. By contrast, most onsite viscometers measure absolute viscosity, but are typically programmed to estimate and report kinematic viscosity, so that the viscosity measurements reported reflect kinematic numbers reported by most labs and lube oil suppliers. Given the importance of viscosity analysis coupled with the increasing popularity of onsite oil analysis instruments used to screen and supplement offsite laboratory oil analysis, it is essential that oil analysts cl